15 Optional Devices
15.5 Setting the CRG Parameters
210
GC-2014
15.5.2
Parameter list
COLUMN OVEN
VALVE
Selection: On/Off, Default: Off
CRG
Selection: Use/Not Use, Default: Not Use
When the CRG is connected to the column oven, the column oven temperature
control range is extended to -99.0
−
+400.0 °C from (room tempe 4 °C)
−
+400.0 °C.
When using the CRG, set “VALVE” to “On” and “CRG” to “Use”.
When using valve control only, set “VALVE” to “On” and set “CRG” to “Not Use”.
INJ 2
VALVE
Selection: On/Off, Default: Off
CRG
Selection: Use/ Not Use, Default: Not Use
When the PTV is installed on INJ2, the injection port temperature control range is
extended to -99.0
−
+400.0 °C from 0.0
−
400.0 °C
When using the CRG, set “VALVE” to “On” and “CRG” to “Use”.
When using valve control only, set “VALVE” to “On” and set “CRG” to “Not Use”.
15.5.3
PF menu
PF menu
Description
Reference
section
Next
Toggles the screen among AOC parameters
→
AUX temperature
→
AUX
APC
→
AUX AMC
→
CRG screens in this order.
――
Содержание GC-2014
Страница 4: ...This page is intentionally left blank ...
Страница 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Страница 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Страница 42: ...2 Before Use 2 6 Connecting the Relay Terminals 30 GC 2014 This page is intentionally left blank ...
Страница 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Страница 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Страница 130: ...11 Creating an Oven Temperature Program 11 2 Temperature Programs 118 GC 2014 This page is intentionally left blank ...
Страница 164: ...12 Injection Port 12 5 Setting the Flow Rate Parameters 152 GC 2014 This page is intentionally left blank ...
Страница 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Страница 318: ...306 GC 2014 This page is intentionally left blank ...