16 Special Functions
16.6 GC Configuration
245
GC-2014
16.6.7.6
Signal offset
When you select [Gain etc.] (PF menu) from the signal setup main screen, the Gain and
offset screen shown in Fig. 16.6.12 appears. Set the offset and the the detector signal
gain here.
Select [Sig. Ch] (PF menu), to set the gain and the offset for any signal other than
defector signals (such as temperature or pressure). In addition, you can set the time
constant and the zero adjustment of the signal for each channel.
Fig. 16.6.13
Detector signal gain and offset screen
Range
Default
Gain offset
0.00-100.00
-999999-9999999
1.00
0
Time constant
4 ms, 5 ms, 10 ms, 20 ms, 50 ms,
100 ms, 200 ms, 500 ms, 1 s, 2 s
50 ms
Содержание GC-2014
Страница 4: ...This page is intentionally left blank ...
Страница 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Страница 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Страница 42: ...2 Before Use 2 6 Connecting the Relay Terminals 30 GC 2014 This page is intentionally left blank ...
Страница 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Страница 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Страница 130: ...11 Creating an Oven Temperature Program 11 2 Temperature Programs 118 GC 2014 This page is intentionally left blank ...
Страница 164: ...12 Injection Port 12 5 Setting the Flow Rate Parameters 152 GC 2014 This page is intentionally left blank ...
Страница 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Страница 318: ...306 GC 2014 This page is intentionally left blank ...