16 Special Functions
16.7 Service and Maintenance
259
GC-2014
TEMP. PORT
Selection: NON/INJ1/ INJ2/DET1/DET2/AUX3/AUX4/AUX5, Default: NON
Specify the injection port heated zone where the flow controller tubing is connected.
This assosiates the flow controller to an injection port.
Q
Detector gas flow controller settings
When installing detector flow control units, the names DET#1 to DET#4 are
automatically assigned starting with the one nearest to the GC.
Specify the to flow controller detector gas settings for each detector No. (DET#1 to
DET#4).
CONT. TYPE
For display only.
The type of each installed detector control unit is automatically recognized and displayed.
DET APC No.
For display only.
When the slot No. is selected for each installed detector gas flow controller, the name
DET APC 1 to DET APC 4 is automatically assigned to each flow controller.
UNIT TYPE
For display only.
When an APC is installed, the APC type is displayed for the Slot No. selected.
“APC (1ch)” indicates an APC for ECD/TCD. “APC (2ch)” indicates an APC for FPD.
“APC (3ch)” indicates an APC for FID/FTD.
When manual flow controllers are selected, “DET GAS” is automatically displayed.
SLOT NO.
Selection: NON/SLOT2
−
6/MSLOT1
−
7, Default: NON
When a manual flow controller is installed, specify the installation location slot no. from
among MSLOT1
−
7.
When a APC is installed, select an available Slot No..
Slot No. which have already been set cannot be selected.
When CONT. TYPE is WDFID (for dual FID) or TCD-L (for packed TCD), two DET APC
numbers and slots are displayed. Only the upper lines can be set. When a manual flow
controller is used, enter the smaller slot no. among the pair (2 slots). (Example: If
MSLOT4 and MSLOT 5 are available, specify MSLOT4.)
Two units of the same type of APCs need to be installed on slots next to each other
because piping for two units is set. An APC for the L side is installed on the slot with the
smaller number so set the slot no. on the L side to the upper line. An APC installed on
the next slot is automatically set on the R side. For example, when the L side is set on
SLOT4, the R side is set on SLOT5.
Setting slots for TCD-L detector gas is unnecessary because makeup gas is usually not
used for packed TCD.
HEATER PORT
Selection: NON/INJ1/INJ2/DET1/DET2/AUX3/AUX4/AUX5, Default: NON
Specify the detector heated zone where the flows controller tubing is connected.
This assosiates the flow controller to a detector.
When CONT. TYPE is WDFID, only heater ports that the DFID is connected to can be
set for piping.
When CONT. TYPE is TCD-L, piping of two units of heater ports need to be set. The
upper line is for the pre heater and the lower line is for the TCD cell. A port that the DFID
(or a pre heater unit without the FID cell) is connected to can be set for the pre heater
and a port that the DTCD is connected to can be set for the TCD cell.
Only heater port that the DFID is connected to is shared for temperature regulation of
the DFID and TCD-L pre heater (set for both).
Содержание GC-2014
Страница 4: ...This page is intentionally left blank ...
Страница 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Страница 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Страница 42: ...2 Before Use 2 6 Connecting the Relay Terminals 30 GC 2014 This page is intentionally left blank ...
Страница 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Страница 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Страница 130: ...11 Creating an Oven Temperature Program 11 2 Temperature Programs 118 GC 2014 This page is intentionally left blank ...
Страница 164: ...12 Injection Port 12 5 Setting the Flow Rate Parameters 152 GC 2014 This page is intentionally left blank ...
Страница 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Страница 318: ...306 GC 2014 This page is intentionally left blank ...