3. Relationship of adjusting plate and oscilloscope output
528µsec
0.1msec/Div
Ø T signal
(PHIT)
Sync signal
Black
1V/Div
1V/Div
Black
White
Black
8lines/mm
4lines/mm
Read line
Read line
Fig. 4
CCD output
Note: The CCD output must be adjusted to produce a white level at
the center and black levels near the center and 8 lines/mm
black and white pattern in other locations. Adjustment should
be made so that the center of the white level signal is delayed
528
µ
sec from the sync signal PHIT.
If this is not obtained the copy image may be shifted left or
right causing incouplete scanning.
AVO
PHIT
CCD signal : AVO
Sync signal : PHIT
GND : DG
DG
Control PWB test pin locations
Fig. 5
FO-4500H
8 – 5