- 16 -
5-2-4. Measure the “DEAD” Weight of Weighing Scale.
When “DEAD” is displayed, press
key, then indicator will calculate Dead weight of scale
part automatically.
Indicator will search “DEAD weight” during 10~20
seconds to find the best condition.
※
To guarantee the preciseness, DEAD weight calculation (CAL00~CAL09) will be operated twice
when resolution
(Division value /Max capacity value)
is less than 1/10,000.
At this step, if there is some force or vibration on weighing scale, and unstable
conditions will be continued,
“Err-A”
will be display, and “DEAD value” will not
be calculated. Please remove all the force or vibration and process it again.
5-2-5. Input Test Weight value and Calculate SPAN value.
If “SPAN” is displayed, input “Test Weight” capacity and press
key.
dead
Cal-00
5paN
1000.0