- 28 -
Steady Range
203
08
01 ~ 99
During the set time period, estimate weighing part’s
“STEADY” condition and display. (Unit: 0.25 gradation)
“STEADY” condition check time
204
10
01 ~ 99
During the set time period, estimate weighing part’s
“STEADY” condition and display. If you set small value, indicator will
take “STEADY” fast, if you set value, indicator will take “STEADY” slow.
(Unit: 0.1 sec)
Digital Filter
205
20
01 ~ 99 01:Weak vibration ~ 99:Strong vibration
Zero key operation
206
●
00
Always active
01
Active under steady condition only
Tare Key operation
207
●
00
Always active
01
Active under steady condition only
Zero key Operation Range
209
00
Active within 2% of Max Capacity
01
Active within 5% of Max Capacity
●
02
Active within 10% of Max Capacity
03
Active within 20% of Max Capacity
04
Active within 50% of Max Capacity
05
Active within 100% of Max Capacity
06
No limit
※
CAUTION: If setting over than 10%, The display weight could be over than Load cell
input signal or Max Capacity and it may display “CELL-Err” or incorrect weight value. And It
can be the cause of load cell damage.
Tare key Operation Range
210
00
Active within 10% of Max Capacity
01
Active within 20% of Max Capacity
●
02
Active within 50% of Max Capacity
03
Active within 100% of Max Capacity
Auto Zero function under Tare state
211
●
00
Disuse
01
Use