- 21 -
If input wrong value, there will display
“Err-01”,
please go back to
Setting “Capacity of Load Cell”.
After recheck the label of load cell and retry the
process.
When “mV” is displayed, input Load cell Rated Output (mV), referring the load cell label. And
press
key to save.
After finishing calculation, calculated “Span
value”, “DONE” will be displayed.
Now, the Simulation Calibration is done, press
key to complete the calibration process.
In case of multiple pieces of load cells are connected, the rated output will be
same as single load cell(Because plural load cells are connected with parallel connection,
the sum of rated output voltage is same as single load cell’s rated output)
※Due to some variation between
“State output rate”
and
“Real Output rate”
of load
cell, there might be some weight difference after finishing calibration. If you want to
make more precise weighing process, please measure real output rate of load cell and
input the measured value. Then the weight measurement will be more precise than
before.