48
SI 480E DIN SIZE WEIGHING CONTROLER
7. Error & Treatment
7-1. Load Cell Installation
Error
Cause
Treatment
Remarks
Weight Value is
unstable
1) Load cell broken
2) Load cell isolation
resistance error
3) Weighing part touches other
devices or some weight is on the
weighing part
4) Summing Board Error
1) Measure input/output
resistance of Load cell.
2) Measure Load cell isolation
resistance
1. Input Resistance of
“EXC+” and “EXC-“ is about
400Ω ±30
2. Output Resistance of
“SIG+“ and “SIG-” is about
350Ω ±3.5
3. Isolate Resistance is
more than 100MΩ
Weight Value is
increased regular
rate, but not return
to “Zero”
1) Load cell Error
2) Load cell connection Error
1) Check Load cell
connection
2) Measure Load cell
Resistance
Weight Value is
increased to under
Zero
Load cell Output wire
(SIG+, SIG-) is switched
Make wire correction
“
”
Load cell broken or Indicator
connection Error
Load cell Check
Load cell connection Check
Power was “ON” when some weight
is on the load cell.
Remove weight
on the Load cell
“
”
(Over Load)
1) Load cell broken or Indicator
connection Error
2) Loading over than Max Capacity
1) Load cell Check
2) Load cell
connection Check
3) Remove over
loaded weight