7-3. Digital Weighing Indicator
Error
No.
Display
Cause
Treatment
No.1
“
CELL-
Er
”
or
“
--OL--
”
1. Load cell Error
2. Load cell cable Error
3.Load cell connection Error
4. A/D Board Error
5.It displays under 5000 or over520000.
1. Under “TEST” mode 1, check analogue
value. If you can not get any analogue value
or there is no change although adding load,
please check load cell, load cell cable,
connection conditions first.
2. Replace another load cell, and check the
indicator condition. If you have same
problem, please replace new indicator and
check A/D board error.
No.2
“
Un-
Pass
”
1. Power is ON, when some materials
are on weighing part.
ଖ
Under “Normal Mode”, if there are
more than 20% loading of Max.
capacity, “
Un-Pass
” display will be
appeared and indicator will stay until
removing the load.
ଖ
Setting Back-up mode it can memory
empty value, and it becomes set value
without displaying ”Un-pass”)
1.If you set “Normal Mode”, please check
weighing part empty or not before turn on the
power. If there are some materials in/on
weighing part, please remove those materials
and turn on the power.
2. Please try to set F02-01(Back-up) mode so
that the indicator can remember first empty
value.
ଖ
Under “UNPASS”, please press
key, then you can exit the mode you
are.
No.3
“FN-
SET”
1. When “FN-Memory” is defected
2. When the “FN-Memory” is empty.
1. Please contact the distributor or Head
Office.
No.4
“P-Err”
Under Parallel Printer is connected and
installed.
1. Parallel printer interface is defected
or disconnected.
1. Please check connection of the print cable.
2. Please check the trouble of print.
ଖ
If you only install “Parallel Print” option
card, you can check to do.
ଖ
Under “CELL-Er”, Relay will not be Output, and Analogue Output(4~20mA/0~10V), either.