Digital Weighing Indicator
SI 4100
G
TGX_GTG
G
6. Define the optimal position or decimal point
Whenever pressing
key, Decimal point will be changed.
pp 30.00
G
7. Press
key to save Digit /Decimal point and
move to next step.
Ex) Load cell CAPA : 30kg, division : 0.01
Input 30.00
p 30.00
G
8. Define the optimal Digit/Division value of weighing measurement.
Whenever pressing
key, the Digit/Division value will
be changed like 1
2
5
10
20
50
Ex) Load cell CAPA : 30kg, division : 0.01
Input division “1”
d 1
G
9. press
key to save the Digit/Division value and
move to next step.
d 1
Caution
ଖ
: (Max. capacity value / division value)
cannot be over
20,000
.(as Indicator resolution
is 1/20,000).
If the value is over 20,000, Error message
“ Err 01 “
will be displayed and move back
“CAPA”
mode again.
10. Under this step, measure the “DEAD Weight of Weighing Scale
When you press
key, the indicator starts to find “Zero”
condition.
dead
G
11. Indicator will search “DEAD weight” during 5sec, automatically.
After finding optimal “Zero” value, automatically move to next step.
Cal-1 0