Digital Weighing Indicator
SI 4100
G
TGZ`GTG
G
Analogue Output Selection (Under 4~20mA Option Card Installed)
F53
ᕢ
0
20mA Output, When Display weight is same as Max. Capacity
1
20mA Output, When Display weight is same as SP1 set value
2
20mA Output, When Display weight is same as SP2 set Value
3
20mA Output, When Display weight is same as SP3 set value
4
20mA Output, When Display weight is same as SP4 set Value
Analogue Output Setting (4~20mA / Option)
F54
ᕢ
0
Positive Output (Max. Capacity : 20mA output)
1
Negative Output (Max. Capacity : 4mA output)
Setting Using Pass word or not (F95 : Set the pass word and change)
F55
ᕢ
0 Using
1
Not
Used
Protocol Frame Transit Setting
F56
ᕢ
0 Not
Used
1
Using (When connecting protocol with an appliance which uses frame by frame.)
Caution : When setting Command frame, if F53(protocol frequency) is high the speed of system can be slow. In
this case, please set F53-01.
BCD INPUT Type Setting (Refer to Interface BCD INPUT)
F57
ᕢ
0
Input the units digit & the tens digit one by one. (1,2,4,8)(1,2,4)
1
Input the units digit & the tens digit together
■ Other Setting
ଖ
Under “Other setting mode”, you can not move to other function directly.
EMPTY Range setting
F80
X.X.X.X.X.X.
(0.0.0.0.1.0)
You can set “EMPTY” Range.
Within set range, indicator will not display current weight and just display “Zero”.
“0.000” setting : When Net Zero, “Zero” status lamp and Near Zero relay will be
output.
“0.190” setting : Within 190, “Zero” Status lamp and Near Zero relay will be output.