- 23 -
SI 400 DIGITAL INDICATOR
209
Zero key Operation Range
02
00: Active within 2% of Max Capacity
01: Active within 5% of Max Capacity
02: Active within 10% of Max Capacity
03: Active within 20% of Max Capacity
04: Active within 50% of Max Capacity
05: Active within 100% of Max Capacity
06:No limit
210
Tare key Operation Range
02
00: Active within 10% of Max Capacity
01: Active within 20% of Max Capacity
02: Active within 50% of Max Capacity
03: Active within 100% of Max Capacity
211
Auto Zero function under Tare state
00
00:Disuse
01:Use
212
Tare Delay Time
00
00:Disuse, 01 ~ 10:Use (Unit:1sec.)
214
Tare reset Time
00
00:Manual (Tare key)
01:Auto (Under near zero range)
02:Auto (At the steady condition)
215
Auto Tare reset Time
00
00 : Disuse
01 ~ 09 : Use (Unit : 1 sec)
216
Hold Mode
00
00:Sample Hold,
01:Peak Hold,
02:Average Hold
217
Hold Delay Time
00
00:Disuse, 01~10:Use (Unit:1sec.)
218
Hold reset at the near zero
00
00:Disuse,
01:Use
219
Auto Hold reset Time
00
00:Disuse, 01~10:Use (Unit:1sec.)
220
Average Hold Time
10
01 ~ 99 (Unit:0.1sec.)
221
Minus (-) Mark Display
00
00:USe
01:Disuse
222
Under UNPASS/OVERLOAD state,
Weight display
00
00:Display,
01:No display
233
External Input 1 Setting
01
00:Disuse, 01:Zero, 02:Tare, 03:Tare reset,
04:Tare/Tare reset 05:Hold, 06:Hold reset,
07:Hold/Hold reset, 08:Print
09:Sub-total print, 10:Grand-total print
234
External Input 2 Setting
04
00:Disuse, 01:Zero, 02:Tare, 03:Tare reset,
04:Tare/Tare reset 05:Hold, 06:Hold reset,
07:Hold/Hold reset, 08:Print
09:Sub-total print, 10:Grand-total print