- 22 -
SI 400 DIGITAL INDICATOR
5-3-2. F-Function List
F-LIST
Subject
Default
Contents
101
Equipment No. setting – ID No.setting
01
01~99
102
Weight–Back up Mode
02
00:Normal mode
01: Weight Back up Mode(Zero)
02: Weight Back up Mode(Zero&Tare)
103
Weighing Data Save Method
00
00: Manual: Whenever “Print” key input
01: Auto: At every steady states
02: Auto: At the first steady states
04: Manual& Auto: At every steady states
05:Manual& Auto: At the first steady
states
104
Display Up-Date Speed
09
01:Slow(1time per 1sec)
~ 09:Fast(60times per 1sec)
105
Main Display setting
00
00 : Weight
01 : Sub-total Weight
02 : Grand-total
108
Buzzer sound (External input detection)
00
00:Buzzer sound, 01:No Buzzer sound
109
Function / Clear key Activation display
selection
00
00 : F or C key press show on the screen.
01 : F or C key press not show on the
screen
110
Weight Unit`
00
00 : kg
01 : g
02 : ton
03 : %
04 : PCS
05 : OZ
06 : lb
111
Language
00
00:Korean, 01:English
201
EMPTY Range
100
00~999999
202
Auto Zero Range
00
00~99 (Unit:0.25gradation)
203
Steady Range
08
01~99 (Unit:0.25gradation)
204
Steady condition check time
10
01~99 (Unit:0.1sec.)
205
Digital Filter
20
01:Weak vibration ~ 99:Strong vibration
206
Zero key operation mode
00
00:Always active
01:Active under steady condition only
207
Tare Key operation mode
00
00:Always active
01:Active under steady condition only
208
Tare key Setting
00
00 : Tare Key
01 : Tare Weight