- 20 -
SI 400 DIGITAL INDICATOR
If input wrong value, there will display “Err-01”, please go back to Setting “Capacity of
Load Cell”.
After recheck the label of load cell and retry the process.
Calculated span value will be displayed. Then
press
key to finish the calibration
step.
In case of multiple pieces of load cells are connected, the rated output will be
same as single load cell’s. (Because plural load cells are connected with parallel
connection, the sum of rated output voltage is same as single load cell’s rated output)
※Due to some variation between
“State output rate”
and
“Real Output rate”
of load
cell, there might be some weight difference after finishing calibration.
If you want to make more precise weighing process, please measure real output rate of
load cell and input the measured value. Then the weight measurement will be more
precise than before.