Plasma Gas Analyzer
15
02001001A_9
4.1
Mechanical installation rack mounting
CAUTION
The analyzer must not be supported by the front panel alone. You must provide an
additional support under the base of the analyzer towards the rear of the enclosure
as shown in Figure 5 - INSTALL THE 02001 ANALYZER IN A PANEL.
If you do not, the analyzer may be damaged, or may fall and damage other
equipment.
Refer to Figure 5 and Figure 6
Prepare a cut-out in a suitable panel capable of supporting 15kg.
Prepare a suitable base support and secure it in your frame or cabinet capable of supporting
15kg.
Fit the analyzer in the panel and use nuts and bolts through the fixing holes in the panel and the
mounting brackets to secure the analyzer in place.
Figure 5 - INSTALL THE 02001 ANALYZER IN A PANEL
Figure 6 - REAR VIEW CUT OUT DIMENSIONS
Содержание SERVOPRO Plasma
Страница 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Страница 2: ...This page intentionally blank...
Страница 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Страница 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Страница 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Страница 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Страница 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Страница 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Страница 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...