Plasma Gas Analyzer
18
02001001A_9
4.2.3 Milliamp output connections
WARNING
The milliamp outputs are separated from the analyzer mains circuits by
reinforced insulation. The terminals must only be connected to circuits
that are themselves separated from mains voltages by an isolation
method that provides at least this level of protection.
CAUTION
To comply with EMC requirements, screened cables must be used to connect
the analogue outputs and the screen of the mA output shall be connected to
the analyzer enclosure. The remote end of the screen shall be open circuit.
Connect the positive and negative wires in your cable to the screw terminals on the rear
panel refer to Figure 7. Section 2.3 provides information on the rating of cable.
The screen should be connected to the SH point on the rear of the analyzer.
Figure 7 - ANALOGUE OUTPUT INTERFACE CONNECTOR
Содержание SERVOPRO Plasma
Страница 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Страница 2: ...This page intentionally blank...
Страница 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Страница 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Страница 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Страница 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Страница 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Страница 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Страница 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...