![Servomex SERVOPRO Plasma Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/servomex/servopro-plasma/servopro-plasma_operators-manual_1247063063.webp)
Plasma Gas Analyzer
61
02001001A_9
1.5
Automatic Calibration Mode
If the 4-20 mA analog output of the analyzer is set to
“HOLD”
mode in the configuration menu,
the analyzer waits for the
“TRANSFER DELAY”
before refreshing the 4-20 mA after the
re-span has been performed. This delay gives the time to purge the span gas out of the
system and avoid undesired bumps on the 4-20mA trending. The analog output remains at the
value calculated just before the beginning of the calibration.
Figure 45 - TRANSFER DELAY
To configure the analyzer to perform automatic calibrations, first set the different timers by
going to the hidden menu. To access the
HIDDEN MENU
, go to the main menu and enter the
secret code by pressing
“1”, “2”, “3”,
and
“E”
consecutively on the keyboard. This brings up
the hidden menu. Press
“F3”
repetitively until the screen of Figure 47 is displayed.
Figure 46 - FIRST PAGE OF THE HIDDEN MENU
<<STARTING COUNT>>
Actual starting count value:25000
Input new value:
F3:NEXT
Содержание SERVOPRO Plasma
Страница 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Страница 2: ...This page intentionally blank...
Страница 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Страница 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Страница 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Страница 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Страница 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Страница 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Страница 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...