Plasma Gas Analyzer
34
02001001A_9
The display shows the actual ppm of nitrogen value and the actual flow in cc. If the unit is started
up for the first time or you replace the calibration gas with a new value, the ppm display has no
meaning. This can be zero or off scale.
In order to calibrate the unit properly, all lines and regulators on the calibration cylinders must
be completely purged. Allow the zero gas to flow into the analyzer. After 10 minutes of the zero
gas flowing into the analyzer, select
"F1"
on the keypad. You will be asked for a confirmation to
re-zero, select
"YES"
, at this moment this message will be replaced by
"PLEASE WAIT"
,
indicating the sampling of the analogue input. This message is displayed for 5 seconds.
NOTE: an indicator is displayed on the right of the confirmation message informing the user if
the re-span or the re-zero has not been performed yet.
After this delay, the
"ZERO DONE"
message is displayed. Now, allow the span gas to flow into
the analyzer for 10 minutes or until the displayed measure value is very stable. Select
"F2"
on
the keypad. Again, you will be asked for a re-span confirmation, select
"YES"
, at this moment,
again, this message will be replaced by
"PLEASE WAIT"
on the display for a certain amount of
time. When this step is done, the
"SPAN DONE"
message is displayed; the span gas value is
shown on the left corner of the display. Recheck zero and, if necessary, calibrate a second time.
When the calibration procedure is finished, go back to the
CALIBRATION MAIN MENU
by
pressing
"F4"
on the keypad. From the
CALIBRATION MAIN MENU
, select
"F4"
again to go
back to the
MAIN MENU
(
TOP LEVEL MENU
).
Содержание SERVOPRO Plasma
Страница 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Страница 2: ...This page intentionally blank...
Страница 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Страница 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Страница 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Страница 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Страница 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Страница 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Страница 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...