Plasma Gas Analyzer
46
02001001A_9
5.8.4 System Gain
This value is used to adjust the last stage of amplification. If for any reason the same amount
of impurity gives less signal, you may adjust this valve to set the amplification. You may see it
as a span potentiometer.
N.B.: You must recalibrate the analyzer each time you change the system gain.
The smallest gain is 2 and the highest is 1000
Minimum: 2
Maximum: 1000
Resolution: 0.01
Figure 36 - SYSTEM GAIN
<<<SYSTEM GAIN>>>
Gain:3.00 Actual:3.00
Signal:87000
Sample flow(CC):75.0
F3:NEXT F4:RET
Detector
G
Analog to
Digital
converter
System
gain