4.1.2 Sample Gas Connections
4.1.2.1 Sample Gas Inlet Connection
Sample gas is connected to the analyzer via a ¼ inch male swivel VCR fitting
labeled Process Inlet at the rear of the instrument as shown in Figure 3. Sample
pressure of 30 – 150 psig is required and is regulated internally.
Pre-purge the line by connecting to the analyzer (with a new VCR
filter
gasket)
only finger tight and flowing gas for 15 minutes. Then tighten the inlet fitting.
NOTE: A VCR
filter
gasket (supplied) should always be used to protect the gas
delivery system from any particulate matter that may have collected in the line.
When power is applied to the analyzer, the internal gas control valves will
automatically go to a state as determined by the user. See the section on Power Up
Defaults on page 60 for additional information.
NOTE: When received from the factory, if equipped with internal gas valves, the
moisture cell will be isolated with pressure in the system. See the sections on
Moisture Gas Valves Control on pages 37 for instructions on starting the gas flow
through either sensor.
See Figure 8 for an overview of the gas flow through the analyzer.
See the section on Gas Pressure and Flow Settings on page 18 for important
information on plumbing and powering up the analyzer.
PROCESS OUTLET
PROCESS INLET
NOT USED
5
5
RLY4-COM
RLY2-COM
RLY1-COM
RLY1-NO
RLY1-NC
KEY
4
4
RLY3-NC
1
2
3
KEY
3
2
1
RLY3-COM
RLY3-NO
RLY2-NO
RLY2-NC
GND
8
6
7
GND
8
7
6
RLY4-NO
RLY4-NC
RX+
TX+
3
3
1
2
1
2
4-20mA-
4-20mA+
TX-
GND
RX-
KEY
8
7
6
4
5
KEY
GND
8
7
6
4
5
AOUT-
AOUT+
~
~
~
~
~
~
Figure 3: Rear Gas Connections
Содержание NanoTrace DF-745SG
Страница 1: ...NanoTrace Moisture Analyzer DF 745SG OPERATOR MANUAL ...
Страница 4: ......
Страница 12: ......
Страница 18: ......
Страница 28: ......
Страница 30: ...24 DF 745SG Options ...
Страница 72: ......
Страница 74: ......
Страница 80: ......
Страница 82: ......
Страница 86: ...80 DF 745SG Appendix A User Menu Screens Page 47 Page 49 Page 52 Page 53 Page 55 Page 56 ...
Страница 87: ...Appendix A User Menu Screens DF 745SG 81 Page 57 Page 58 Page 59 Page 60 Page 61 Page 62 Page 63 Page 64 ...
Страница 88: ...82 DF 745SG Appendix A User Menu Screens ...
Страница 93: ......