10 DF-745SG
Specifications
Sample Temperature:
10° C to 80°C (50° F to 176° F)
EMI Sensitivity:
Tested to standards EN61000-3-3 and EN61326-1
Audible/Visual Alarm Status Indicators:
Four Moisture levels,
Temperature, Moisture cell Diagnostic, Zero Verification or Calibration in
Process, Moisture Analyzer off-line, Analog output freeze during calibration.
Alarm Relays:
Four non-latching, independently assignable to moisture alarms
or to moisture calibration-in-process indicator. SPDT contacts rated at 1 Amp @
30 VDC. Fail safe action upon loss of power to alarm condition. Not designed to
switch AC power.
Storage Temperature:
Not to exceed 50° C (122° F)
Construction:
NEMA 1, 19 inch rack mount
Dimensions:
19 inch (48.3cm) wide x 10.5 inch (26.7 cm) high x 22.5 inch (57.2
cm) deep
Weight:
70 pounds (31.8 kg)
Gas Path Construction Materials:
300 series stainless steel electro-polished
1/4-inch
VCR
-type compatible sample inlet fitting
1/8-inch compression sample outlet fitting
1/4-inch compression vacuum fitting
Kel-f valve seats
PTFE Vacuum tubing
Содержание NanoTrace DF-745SG
Страница 1: ...NanoTrace Moisture Analyzer DF 745SG OPERATOR MANUAL ...
Страница 4: ......
Страница 12: ......
Страница 18: ......
Страница 28: ......
Страница 30: ...24 DF 745SG Options ...
Страница 72: ......
Страница 74: ......
Страница 80: ......
Страница 82: ......
Страница 86: ...80 DF 745SG Appendix A User Menu Screens Page 47 Page 49 Page 52 Page 53 Page 55 Page 56 ...
Страница 87: ...Appendix A User Menu Screens DF 745SG 81 Page 57 Page 58 Page 59 Page 60 Page 61 Page 62 Page 63 Page 64 ...
Страница 88: ...82 DF 745SG Appendix A User Menu Screens ...
Страница 93: ......