36 DF-745SG
User Interface
top of the supported gases list, at which point the arrow on the screen
will turn red. Hit the
Up
arrow key one more time and the screens will
toggle between Supported Gases and Template. Note that the red
arrow key is displayed only when the top item is selected in the
"Supported Gases" or "Templates" screen.
3. Use the
Up
and
Down arrow
keys to highlight a template from the list.
4. Use the
Right arrow
key to select a particular template which
populates the Sample Gas Mixture table in the Supported Gases
display.
5. Edit the contents of the mixture as described in Section 7.4.1.1.1 and in
Section 7.4.1.1.2.
6. Rename the template if desired as described in Section 7.4.1.2.1.1.
7. To use the edited template, but not rename it, navigate with the
Next
key to the
Accept
button and hit enter.
7.4.1.2.4 Deleting an existing template
1) The GSF Template screen displays available user-defined templates (up to
30 are allowed). Use the
Up
and
Down arrow
keys to highlight a template
from the list and to display the recipe (gases, percentages) in the table to
the right of the template list.
2) Use the
Left arrow
to delete a template from the table. Deletion of a
template prompts a confirmation.
7.4.2 Isolate Analyzer – Optional
Figure 19: Isolate Analyzer
Isolate Analyzer allows the internal gas lines and the moisture sensing cell volume
to be isolated in the case of an impending break in the external delivery lines. It is
also considered the first step in the shut down procedure. Highlight this item and
hit the
Enter
key to initiate the isolation process. The Isolate Warning shown in
Figure 20 will appear for 40 seconds.
Содержание NanoTrace DF-745SG
Страница 1: ...NanoTrace Moisture Analyzer DF 745SG OPERATOR MANUAL ...
Страница 4: ......
Страница 12: ......
Страница 18: ......
Страница 28: ......
Страница 30: ...24 DF 745SG Options ...
Страница 72: ......
Страница 74: ......
Страница 80: ......
Страница 82: ......
Страница 86: ...80 DF 745SG Appendix A User Menu Screens Page 47 Page 49 Page 52 Page 53 Page 55 Page 56 ...
Страница 87: ...Appendix A User Menu Screens DF 745SG 81 Page 57 Page 58 Page 59 Page 60 Page 61 Page 62 Page 63 Page 64 ...
Страница 88: ...82 DF 745SG Appendix A User Menu Screens ...
Страница 93: ......