PREVENTIVE MAINTENANCE & REPAIR
Preventive Maintenance
The recommended periodic preventative
maintenance is as follows:
Perform Quarterly Filter Inspection and
replace as necessary.
Perform Quarterly Performance Verification.
Filter Inspection and Replacement
Coarse Filters (Tube End Filters)
are
located in each sample zone and fresh air
zone on the ends of the 1/4 inch tubes.
The Coarse Filters are sintered brass or
bronze, and will naturally discolor to a dark
bronze color. Inspect the filter for obvious
clogging or build up of dust. Replace,
rather than clean the filters.
5 Micron Filter
is located inside the
monitor, inside a clear moisture separator.
The 5 micron filter should be ‘paper white’,
and will turn gray, or off color when dirty.
Replace, rather than clean this filter.
Quarterly Performance Check:
Once per
quarter, perform the End-to-End system
test described in the Final Checks
Procedure on page 12.
Commonly Ordered Parts
P/N
Description
410144
Filter,
Coarse Tube
410162
Filter,
5 Micron
410136-GRAY
Orifice, .016
410224
Diaphragm
Pump
Monitor Repair
Most monitor repairs can be performed in
the field by the typical service technician.
Refer to the Troubleshooting Guide found
at the end of this manual for assistance.
Additionally, SenTech Corporation provides
telephone technical support for all SenTech
refrigerant monitors. If assistance is
needed, call 888-248-1988 (317-248-1988
internationally).
Should a repair be beyond the normal
scope of field service, contact SenTech
Corporation for a Return Authorization
number for factory repair.