4. Pneumatics Failure
5. Failure to Detect a Leak or
Suspected Low Sensitivity
power fault in the control board, or a wiring short causing
a power fault in the control board. Turn off the monitor
and remove one connector at a time from the control
board. Apply power between each step. If the dc
voltages remain out of spec, the control board has failed.
a) Check that the diaphragm pump is operating and
replace if it has failed.
b) Check that the tubing inside the monitor has not come
loose from any fitting, or become kinked at any point.
c) Check that the orifice or tube end filter have not
become clogged or restricted.
d) Check that the tubes connected to the sample air inlet
and fresh air inlet have not become clogged or restricted.
e) Check that the solenoid valves are switching. Verify
that the light labeled “RLY6” is switching on and off.
Verify that the red LED’s on turning on an off, according
to which area is being sampled. Verify that the voltages
on Valve Driver Board, J4 are changing from
approximately 24 vdc to approximately 0 vdc, according
to which area is being sampled. Temporarily block the
air inlet and listen for the diaphragm pump to draw a
vacuum. If the pump does not change tone, and is
known to work, the valve is not switching.
a) Alarm lights are on, but external alarm devices are not
actuated. Repeat the “Integration with External Devices”
from the Final test portion of this manual.
b) Alarm light is not on, but a leak is suspected. Verify
that the alarm settings are not set too high.
Refer to “Startup and Programming” section of this
manual for instructions.
c) Repeat the complete procedures described in the
“Final Tests” section of this manual.
d) Verify that the sample air and fresh air pick up point
locations are appropriate in accordance with the
“Installation” section of this manual.