The calibration of the temperature channel is equivalent to the oxygen calibration.
Under the item Information you can see serial number of the sensor (SN) the signal at the
lower calibration point (C0) and upper calibration point (C1) as well as the sensor mode.
By selecting factory settings, you can insert calibration values manually.
Alarm thresholds
In case there should be an alarm sent, when the sensor gives readings above or below a certain
threshold, you can define these thresholds in this menu item. The hysteresis defines the
tolerance of alarming. For example, if the lower threshold of 70% defines the alarm level and
the concentration fluctuates between 69 and 71%, the hysteresis of 3% eliminates several
warnings always when the sensor reaches 70%. With alarm on / off you can also deactivate
the alarming for this sensor.
Plug name
With plug name you can give the sensor a name of max. 8 characters, which will be displayed
in the SENECT Control App.
Salinity correction
Since the oxygen saturation concentration depends on the salinity of the water, the SENECT
control can correct for the effect of salinity. Therefore, you can either enter a specific salinity
(Mode: Correction value) or connect a conductivity sensor (Mode: Sensor corrected), which
calculates the salinity and the correction based on conductivity measurements.
Menu output ports and 230 V AC plugs
Output / 230 V plugs
With the AQUACULTURE|CONTROL up to 6 actuator output ports (OUT 1 to OUT 3 / OUT 6)
and three mains plugs (230 V OUT 1 to 230 V OUT 3) can be indivdually configured to fulfill
specific functions.
The actuator output ports are designed for the use of low-voltage consumer with up to 24 V
DC or to control actuators with control signals (4-20 mA or 0-10 V).
At the end of each line, the current state is displayed:
Output port switched on
Output port switched off