10 Error messages
Error 1
Pressure sensor failure or overpressure.
Error 2
PT100 probe failure or overtemp.
Error 3
External alarm. Pressure switch or safety thermostat trigered.
To reset the safety thermostat, press its button num-
ber 20.
The thermostat is adjusted in the factory, it must not be
operated except to reset it.
Error 4
Open door interlocking during the cycle.
Error 5
Open door during a cycle.
Error 6
Pressure over +0.05bar with the autoclave stopped.
Error 7
E2PROM failure.
Error 8
Out of limits temperature during sterilization.
Error 9
Mains supply failure during cycle.
11 Spare parts
07292 RS-232 communication wire
07315 26 pins communication wire
10044 R4 closing
13006 Contactor
15410 Fuse 16 A
16055 Solenoid electromagnet
16057 Electrovalve NA 3 mm 24 V
16212 Pressure switch
20096 Bipolar switch
29358 Frontal circuit
29459 Power circuit
36027 Solid relay 25A
39001 Heating element 2800W/230V
43031 Thermostat
43071 Pt100 probe
12 Lid emergency opening
In case of power supply failure, it is possible to open the lid manually.
Attention: Before opening the cover by using this system, make
sure that the gauge pressure is 0 bar.
Remove the top drawer from the display by pulling hard (it is not screwed
down, but snapped). A yellow cap will appear, gently pull it until you
hear a click. The closure will be opened.
20
13 Recommendation to achieve a perfect
sterilization
The material to be sterilized needs to be clean, free from any encrusted
or residual material. We recommend to wash it with a good detergent
and distilled water, then rinse it with distilled water.
Do not overload the baskets, drums or trays, always leave a space
of 1 or 2 cm between items to allow a free passage of vapour and to
ease drying.