9.2 Safety thermostat
8. Safety thermostat regulation screw (grey).
9. Safety thermostat reset button (white).
9.2.1 Rearming the safety thermostat
Safety thermostat is an element to prevent the temperature from exceeding
a value. If by a fortuitous event the thermostat is triggered, that is, it opens
the power circuit, the heating element will stop working. Usually there is an
indicator light, or an error message, that indicates that the thermostat is in this
position.
If the thermostat is triggered, check that the equipment looks good and it is not
overheating at any point.
To change its status, simply press the white button (9), and you will hear a
“click” that indicates that the thermostat has been reset. If the equipment has
not cooled down sufficiently, the thermostat cannot be reset, and you must wait
for the temperature to drop before it can be reset.
9.2.2 Adjusting the safety thermostat
In some equipment such as ovens or baths and with certain processes, it may
be useful to adjust the thermostat trigger value, but for the rest of equipments
this operation should NEVER be carried out.
From factory, the thermostat is adjusted to its maximum temperature.
9.3 Gasket
It is necessary to check that the gasket is in good condition. With the door
open, remove the gasket and check that it is not deformed, has cracks or is
broken. To check it, remove the gasket from the cover, dry it with a cloth and
dry the groove of the cover where the gasket is located. Reassemble and do a
test cycle. To ensure a good seal with a use of 1 cycle a day, it is recommended
to replace the gasket every two years.
9.4 Cleaning the chamber
When working with steam, it is possible that traces of materials are deposited
on the walls of the chamber. It is recommended from time to time to clean the
inside of the chamber with a clean cloth. Also remove the heater cover and
clean the bottom of the chamber.