7 Installation
1 Locate the autoclave close to an appropriate power source.
2 The autoclave needs to be placed on a flat stable horizontal sur
-
face, leaving a minimum space of 10 cm around the instrument.
3 Adjust the front feet to level the autoclave and set the autoclave
in position.
4 Fix a plastic or rubber hose to the pump output connection (21)
with a receptical to collect any condensate produced by the at
-
mospheric purge.
You must ensure that the purge condensate tube is not
clogged or submerged in water after the cycle, since the
chamber will empty and you will not be able to open the
lid.
Note that during the purge phase, the equipment draws steam
through this tube and this steam is in form of water inside the
container. Make sure the tube is not in contact with the bottom
of the container to ensure that at the end of the cycle it does not
absorb water from it.
18. Safety valve
19. Power supply cord
20. Thermostat reset button
21. Condenser outlet and chamber emptying
21
18
19
20