- End cicle: When sterilization is finished, cold process starts. The lamp
(5) is blinking and the lamp (3) lights on. The display (6) shows the
internal real temperature.
In case of solids sterilization, fast steam output can be made using
button (14), the steam output is made with output (21).
If liquids or culture media are sterilized, you must leave it cool until
the pressure will be 0 bar. During the cooling, it is the same case of
manual depressurize, the display shows the internal temperature.
When the pressure is 0, the temperature is under 95ºC, the display
(6) shows the word «End». Therefore, you can open the lid with its
handle (16).
YOU CAN NOT OPEN THE LID UNTIL THE PRESSURE IS NOT
0 BAR AND THE INTERNAL TEMPERATURE UNDER 95ºC.
SEE SECTION 8.1
DO NOT OPEN THE VALVE 16 UNTIL
THE CYCLE IS COMPLETELY FINISHED TO
AVOID POSSIBLE BURNS
WITH HEAT WATER OR STEAM.
¡WARNING!
- THE DIRECT CONTACT WITH PRESSURE
STEAM CAN PRODUCE SERIOUS BURNS.
- IF YOU USE A HOSE TO MAKE THE STEAM
OUTPUT, DON NOT HOLD IT WITH HANDS
TO AVOID BURNS.
When the lid is ready to be opened, the display (6) shows the word
«End» and the lamp (5) is off.
6
9 Maintenance
9.1 Temperature calibration
If a correction needs to be done, do as follows:
- Press the button (8) «ºC» and (9) «clock» simultaneously for more
than 5 seconds.
- While continuing to press these two keys, press keys (10) and (11)
to modify this parameter. For fast forward press and hold the arrow
(10 or 11) for a few seconds.
Note: Correction range from -9.9 to + 9.9ºC. The modified value is
added or restored to the measured temperature by correcting it.
9
8
10
11