Messages in the 'Status' line:
Saving
the instrument is saving the performed calibration
Error
the sensor has NOT been recalibrated for any of the following reasons:
- The calibration gas cannot properly reach the instrument.
- Concentration for the calibration gas has not been set in the relevant line ‘Applied gas’.
- The user didn’t allow for the stabilization time to properly elapse.
- The sensor could be damaged or exhausted and must therefore be replaced.
When the stabilization time is over, select 'Calibration' and store the new calibration.
11/12/17 10:00
ON SITE CALIB.
►
Calibration
Status ----
Durat. 00:00:00
Appl 800 ppm
Meas 800 ppm
Is 0.52 uA
11/12/17 10:00
ON SITE CALIB.
►
Calibration
Status ON
Durat. 00:00:00
Appl 800 ppm
Meas 800 ppm
Is 0.52 uA
OK
WARNING!
At any time the user can restore the factory calibration in the instrument by setting the
'Status' line on 'OFF'.
Below are listed the suggested stabilization times for the ‘on site calibration’ of the
sensors:
CO sensor:
3 minutes
NO sensor:
3 minutes
11/12/17 10:00
ON SITE CALIB.
Calibration
Status ----
►
Durat. 00:03:60
Appl 800 ppm
Meas 0 ppm
Is 0.52 uA
Resets the timer (helps to control
the elapsed time during the
stabilization phase)
OK
The instrument measures the concentration of gas applied;
wait at least 3 minutes to allow the reading to
stabilize
. The reading is shown in line 'Meas'.
11/12/17 10:00
ON SITE CALIB.
Calibration
Status ----
►
Durat. 00:00:00
Appl 800 ppm
Meas 0 ppm
Is 0.52 uA