To perform complete flue gas analysis, follow the instructions below.
SOME IMPORTANT WARNINGS TO CONSIDER DURING THE COMBUSTION ANALYSIS
ARE LISTED BELOW:
FOR A CORRECT ANALYSIS NO AIR MUST FLOW INTO THE PIPE FROM OUTSIDE DUE
TO A BAD TIGHTENING OF THE POSITIONING CONE OR A LEAK IN THE PIPELINE.
THE GAS PIPE MUST BE CHECKED IN ORDER TO AVOID ANY LEAKAGES OR OBSTRUCTIONS ALONG
THE PATH.
THE CONNECTORS OF THE GAS SAMPLING PROBE AND OF THE CONDENSATION FILTER
MUST BE WELL CONNECTED TO THE INSTRUMENT.
KEEP THE CONDENSATION TRAP ON VERTICAL
POSITION DURING THE ANALYSIS; A WRONG POSITIONING MAY CAUSE CONDENSATE
INFILTRATIONS IN THE INSTRUMENT AND THUS DAMAGE THE SENSORS.
AFTER EACH ANALYSIS
CHECK FOR ANY PRESENCE OF WATER IN THE CONDENSATE COLLECTION BOWL AND REMOVE IT
IF ANY. PUT THE PROBE BACK IN THE CASE ONLY AFTER YOU HAVE ELIMINATED CONDENSATE
FROM THE TUBE AND THE EXPANSION TANK (SEE CHAPTER '
').
REPLACE THE FINE
DUST FILTER IF IT IS VISIBLY DIRTY OR WET (SEE CHAPTER '
'). DO NOT PERFORM
ANY MEASUREMENT WHEN THE FILTER IS REMOVED OR DIRTY IN ORDER TO AVOID ANY RISK OF
IRREVERSIBLE DAMAGES TO SENSORS AND ANALYZER ITSELF.
8.0 FLUE GAS ANALYSIS
8.1 FLUE GAS ANALYSIS
11/12/17 10:00
AUTOZERO
Primary air
acquisition
T: 21.5°C
keep
Repeat
11/12/17 10:00
AUTOZERO
60
Primary air
acquisition
Keep
pressed for
a few sec-
onds
S500-2
N.S.:0
Fw:1.00-EVAL03
BEFORE TURNING ON THE INSTRUMENT:
CONNECT THE GAS SAMPLING PROBE TO THE INSTRUMENT.
STORING OF THE AMBIENT TEMPERATURE: UPON COMPLETION OF THE AUTOZERO
IN FRESH OUTDOOR AIR, PRESS THE “KEEP” BUTTON TO STORE THE OUTSIDE
TEMPERATURE BEING USED FOR PRIMARY AIR.
IF THE Tc-K CONNECTOR IS NOT PLUGGED IN, THE TEMPERATURE WILL NOT BE
ACQUIRED.
8.1.1 Switching on the instrument and auto-calibration
11/12/17 10:00
HARDWARE
►Memories
KO
Calibration KO
voltages
ADC channels
WARNING!
When the instrument is turned on, an hardware check
is performed on the memories and on the instrument
calibration. If some error occurs, they will be shown
with the activation of the
Hardware
screen. In this
case, it is advisable to restart the instrument. If the
problem keeps on appearing or frequently occurring
please contact the Seitron Service Center,
communicating the shown error.