To perform the recalibration the following instruments are needed:
- Known concentration gas cylinder suitable for the sensor, equipped with a pressure regulator
- Flow meter
- Hose with T-fitting to connect the cylinder to the flowmeter and to the instrument
WARNING!
For the oxygen sensor on site calibration, the zero value calibration must be carried out with nitrogen or
any other gas mixture which DOES NOT contain oxygen.
1. Start the instrument
WARNING
Be sure that the autozero is performed in clean fresh air and that it terminates
correctly.
Do not connect the gas probe to the instrument.
Check the battery charge level or connect the power adapter to avoid data loss
during recalibration.
On site calibration procedure
2. Once the autozero is completed press the button and select the menu
Configuration
→
Instrument
→
on site calib.
Keep
pressed for
a few
seconds
11/12/17 10:00
AUTOZERO
60
Primary air
acquisition
S500-2
N.S.:0
Fw:1.00-EVAL03
11/12/17 10:00
AUTOZERO
Primary air
acquisition
T: 21.5°C
Keep
Repeat
11/12/17 10:00
ON SITE CALIB.
1110
1
Insert password
' 1111 '
>
>
>
>
OK
OK
11/12/17 10:00
ON SITE CALIB.
000
0
11/12/17 10:00
INSTRUMENT
Display
►
On site calib.
Clock
Bluetooth