43
K100000000S9 036103 130121
12.3 Menu→Configuration→Instrument
SUB MENU
FUNCTION
Display
With the arrow keys it is possible to increase or decrease the brightness and the
contrast of the display.
On site calib.
It is possible to make a recalibration of the instrument’s gas sensors with suita-
ble known concentration gas cylinders. Recalibration of Oxygen (O
2
) sensor is
not available since it is already recalibrated during every autozero sequence.
The access to the sensor recalibration is password protected, the pass-
word is ' 1111 '.
Clock
This allows the current time and date to be set. The user can select the date and
hour format either in EU (European) or USA (American) mode.
Bluetooth
(if the version of
the instrument
foresees it)
In this sub menu it is possible to turn on and off the Bluetooth
®
communication
of the instrument and to visualize the related codes.
15/06/18 10:00
INSTRUMENT
►
Display
On site calib.
Clock
Bluetooth
KEY
FUNCTION
Returns to the previous screen.
Selects the available parameters.
Enters in the selected parameter.
ESC
>
>
OK