10.3 Menu→Measurements→Draft
15/06/18 10:00
DRAFT
Inlet P+
Draft 0.00 h
►Zero sensor
Save
To measure the draft proceed as follows:
- Connect the probe pressure input hose to the instrument
P+
input.
- Before starting the pressure zeroing sequence pay attention to remove the gas probe from the
stack.
- Having carried out the pressure zeroing sequence, insert the probe in the stack and measure the
draft.
INTERACTIVE OPERATION FUNCTION
Save
The measure will be printed on the ticket of the current combustion analysis
According to the version of the instrument and/or according with the related
setting, it is possible to print or visualize the QR code
KEY
FUNCTION
Activate the context keys shown on the
display.
Returns to the previous screen.
Starts the pressure sensor autozero.
>
>