6.0 TECHNICAL SPECIFICATIONS
6.1 Technical specifications
Power:
Li-Ion battery pack with internal protection circuit.
Average life of the battery package: 500 empty / full charge cycles.
Battery charger:
External 5Vdc 2A battery charger with female A-type USB connector +
connection to the device with the same serial communication cable
supplied.
Charging time:
5 hours to charge from 0% to 90% (6 hours for 100%). The device can also
be charged by connecting it to the PC, the device must be turned off, the
charging time depends on the output current from the PC and may be more
than 12 hours.
Instrument working time:
8 hours of non-stop operation with display brightness at 50%.
——————————————————————————————————————————————————
Display:
Graphic white LED backlit White / Black, 128 x 128 pixel
——————————————————————————————————————————————————
Connectivity:
Communication port:
USB connector type B.
Bluetooth
®
(if the version of the
instrument foresees it):
Class 1. Communication distance <100 meters (in open field)
Infrared interface:
For external printer (optional) using protocol HP-IR.
——————————————————————————————————————————————————
Autozero:
Settable (30 .. 600 seconds)
——————————————————————————————————————————————————
Gas measurement sensors:
Up to 3 electrochemical sensors
Type of combustible:
12 predefined by the factory.
————————————————————————————————————————–——————--——--
Self-diagnosis:
Checks all functions and internal sensors and reports any abnormal
operation.
Temperature measurement:
Input for thermocouple type K with mignon connector (ASTM E 1684-96) for the
temperature measurement.
Room temperature measurement: Through the internal sensor and/or acquisition through the gas probe
positioned in air.
——————————————————————————————————————————————————
Internal data memory:
5 complete analyses.
——————————————————————————————————————————————————
Suction pump:
0.26 gal. per minute (1.0 l/min) heads at the flue up to 80hPa.
——————————————————————————————————————————————————
Condensate trap:
Type:
Outside the instrument.
Line filter:
With replaceable cartridge, 99% efficient with 20um particles.
——————————————————————————————————————————————————
Condensing boiler efficiency:
Automatic recognition of the condensing boiler, with calculation and printout
of efficiency (>100%) on the LHV (Lower Heating Value).
Environmental gases:
Measurement and separate printout of the ambient CO values.
Draft test:
By using the internal sensor connected to the port P+.
——————————————————————————————————————————————————
Operating temperature range:
23°F .. 113°F (-5°C .. +45°C)
Storage temperature range:
-4°F .. 122°F (-20°C .. +50°C)
Humidity limit:
20% .. 80% RH
Protection grade:
IP42
Air pressure:
Atmospheric
Outer dimensions:
Analyzer:
2,8 x 2,4 x 6,7 inches (Width x Depth x Height)
Case:
15,7 x 11,4 x 4,7 inches (Width x Depth x Height)
Weight:
Analyzer:
~ 12,3 Oz
——————————————————————————————————————————————————
See the declaration of conformity (ANNEX E).