Manual revision 041
Section 4: Deploying and Operating MicroCAT
SBE 37-SM RS-232
31
Status
Commands (
continued
)
GetSD
Get and display status data, which contains
data that changes while deployed.
List below includes, where applicable,
command used to modify parameter:
Device type, Serial number
Date and time [
DateTime=
] in
ISO8601-2000 extended format
(yyyy – mm-ddThh:mm:ss)
Number of recorded events in event
counter [reset with
ResetEC
]
Voltages – main battery pack voltage and
back-up lithium cell voltage
Memory – [reset with
InitLogging
]
- Number of bytes in memory
- Number of samples in memory
- Number of additional samples that can
be placed in memory
- Length (number of bytes) of each sample
Logging status –
yes or no (to indicate whether it is
currently logging data);
if applicable, reason that logging has
stopped
Example:
(user input in bold, command used to modify parameter in parentheses)
getsd
<StatusData DeviceType = 'SBE37SM-RS232' SerialNumber = '03730051'>
<DateTime>2017-08-20T00:48:32</DateTime>
[
DateTime=
]
<EventSummary numEvents = '0'/>
[can clear with
ResetEC=
]
<Power>
<vMain>13.58</vMain>
<vLith> 2.93</vLith>
</Power>
<MemorySummary>
<Bytes>0</Bytes>
<Samples>0</Samples>
[can clear with
InitLogging
]
<SamplesFree>559240</SamplesFree>
[can clear with
InitLogging
]
<SampleLength>15</SampleLength>
</MemorySummary>
<AutonomousSampling>no, stop command</AutonomousSampling>
[
StartNow
or
StartLater
,
Stop
]
</StatusData>