![Sea-Bird Scientific SBE 37-SM Скачать руководство пользователя страница 48](http://html.mh-extra.com/html/sea-bird-scientific/sbe-37-sm/sbe-37-sm_product-manual_1221402048.webp)
Manual revision 041
Section 4: Deploying and Operating MicroCAT
SBE 37-SM RS-232
48
OutputFormat=3
: converted decimal data, alternate format
ttt.tttt,cc.ccccc, pppp.ppp, sss.ssss, vvvv.vvv, dd mmm yyyy, hh:mm:ss
where
o
ttt.tttt = temperature (sent if
OutputTemp=Y
; units defined by
SetTempUnits=
).
o
cc.ccccc = conductivity (sent if
OutputCond=Y
; units defined by
SetCondUnits=
).
cc.ccccc if
SetCondUnits=0
(S/m)
cc.cccc if
SetCondUnits=1
(mS/cm)
ccccc.c if
SetCondUnits=2
(µS/cm)
o
pppp.ppp = pressure (sent if optional pressure sensor installed and
OutputPress=Y
; units defined by
SetPressUnits=
.)
Number of digits to left of decimal place is dependent on pressure
sensor range.
o
sss.ssss = salinity (psu); sent if
OutputSal=Y
.
o
vvvv.vvv = sound velocity (m/sec); sent if
OutputSV=Y
.
o
dd mmm yyyy = day, month, year.
o
hh:mm:ss = hour, minute, second.
Leading zeros are suppressed, except for one zero to the left of the
decimal point. All data is separated with a comma and at least one space.
Example:
Sample data output for real-time autonomous data when pressure sensor is installed,
OutputFormat=1
,
SetTempUnits=0
,
SetCondUnits=0
,
SetPressUnits=0
.
Spaces in this example have been replaced with a · symbol to
better demonstrate the data string
:
21.5055,··0.0000,···0.742,·18·Oct·2018,·15:31:54
(temperature, conductivity, pressure, date, time)
Notes:
OutputFormat=3
matches the
output from SBE 37-SMs (firmware
version 3.1 and below).