Section 4: Deploying and Operating MicroCAT
SBE 37-SM RS-232
35
Status
Commands (
continued
)
DS
Display operating status and setup.
List below includes, where applicable,
command used to modify parameter.
Firmware version, serial number, date and time
[
DateTime=
].
Main battery pack voltage and back-up lithium
cell voltage.
Number of samples in memory
[
SampleNumber=
] and available sample space
in memory.
Logging status (logging not started, logging
data, not logging, or unknown).
Interval between samples for autonomous
sampling [
SampleInterval=
].
Output temperature [
OutputTemp=
]?
Temperature units [
SetTempUnits=
]
Output conductivity [
OutputCond=
]?
Conductivity and specific conductivity units
[
SetCondUnits=
]
Output pressure [
OutputPress=
]? Pressure
units [
SetPressUnits=
]
Output salinity [
OutputSal=
]?
Factory-set salinity units (psu)
Output sound velocity [
OutputSV=
]?
Factory-set sound velocity units (m/s)
Output specific conductivity [
OutputSC=
]?
Conductivity and specific conductivity units
[
SetCondUnits=
]
Specific conductivity temperature coefficient
[
UseSCDefault=
and
SetSCA=
]
Transmit sample number with real-time
autonomous data and polled data from memory
[
TxSampleNum=
]?
Transmit autonomous and serial line sync data
real-time [
TxRealTime=
]?
Serial sync mode state [
SyncMode=
].
Reference pressure to use in calculations if no
pressure sensor installed (only sent if pressure
sensor not installed) [
ReferencePressure=
].
Example:
MicroCAT with a pressure sensor (user input in bold, command used to modify parameter in parentheses).
DS
SBE37SM-RS232 v5.0.0 SERIAL NO. 30051 04 Aug 2017 20:55:10
[
DateTime=
]
vMain = 13.59, vLith = 2.91
samplenumber = 3418, free = 559240
[
SampleNumber=
]
not logging, stop command
sample interval = 60 seconds
[
SampleInterval=
]
data format = converted engineering
[
OutputFormat=
]
output temperature, Celsius
[
OutputTemp=, SetTempUnits=
]
output conductivity, S/m
[
OutputCond=, SetCondUnits=
]
output pressure, Decibar
[
OutputPress=, SetPressUnits=
]
output salinity, PSU
[
OutputSal=
, factory-set units]
output sound velocity, m/s
[
OutputSV=
, factory-set units]
output specific conductivity, S/m
[
OutputSC=
,
SetCondUnits=
]
specific conductivity coefficient = 0.0200
[
UseSCDefault=
and
SetSCA=
]
output sample number
[
TxSampleNum=
]
transmit real time data = yes
[
TxRealTime=
]
sync mode = no
[
SyncMode=
]
Notes:
The
DS
response contains similar
information as the combined
responses from
GetSD
and
GetCD
, but in a different format.
Lines describing what parameters
to output (temperature,
conductivity, pressure, salinity,
sound velocity, specific
conductivity, sample number) only
appear if they are enabled, and if
OutputFormat=1
or
2
. Raw output
(
OutputFormat=0
) is not affected
by enabling / disabling parameter
outputs.
The
DS
response is also affected
by the
Legacy=
command. See the
Output Format Setup commands
below.