LUBExpert User Manual
20/52
7.2.3.
Assigning LUBExpert Sensors
If you followed the instructions to go into LUBExpert Mode, the
list of available sensors should only be two (LUBEsense1 and
TEMP2). The other sensors are still visible but inactive. Choose
LUBEsense1
to create an ultrasound measurement node and
TEMP2
to create a temperature measurement node as required.
For ultrasound measurements, UAS3 gives you the option to
choose either Static or Dynamic and for temperature
measurements you can set the measurement interval and
the emissivity.
If your instrument has only Static capabilities (LUBExpert,
SDT270 SU), you should choose Static. If Dynamic is
selected, your instrument will not recognize the
measurement.
However, if you are using an SDT270DU with the LUBExpert
App installed or LUBExpert Dynamic, choosing Dynamic here will instruct your instrument to collect
both Dynamic and Static data simultaneously.
7.2.4.
Setting Data Collection Interval, Acquisition Time, and Temperature Emissivity
Setting the interval of data collection
For both LUBEsense1 and TEMP2
measurements, the choice to set your
preferred data collection interval time
①
in
months, weeks, days, or hours
②
is made
here.
Note that this field is not intended to be the
re-greasing interval. This field is your data
collection interval.
Condition assessment based on measured
ultrasound data will be the trigger to
perform grease replenishment (or not) as
well as the quantity of grease required to
restore an optimal lubrication mechanism to
the bearing.
For TEMP2 measurements the additional
choice to set emissivity exists here. The
default emissivity is 1.0.
Setting Grease Type
Earlier we described how to set up a list of
all the lubricant types used in your plant
under “Lubricant Management”. With the
“Create New Measurement” dialogue box open, here you select the grease type for this bearing
using the drop-down box beside the field Grease Name
③
.
②
①
③
④