Figure C-2 Connecting the Sample Line to the Union
3. Insert the fitting into the upstream side of the union, pushing the tubing into the fitting as far as it will go.
4. While holding the tubing in place, tighten the fitting slightly.
5. Insert the other end of the red PEEK-line fused silica tubing through the stainless steel nut and ferrule, making
sure that it protrudes about 2 mm.
6. Insert the PEEK-line fused silica tubing, with the nut and ferrule, into the syringe union, pushing the tubing in
as far as it will go, and then use a 1/4 inch and a 3/16 inch wrench to tighten the nut.
Figure C-3 Syringe Union, Ion Source Side
7. Cut a 3 cm length of the green PEEK sleeve.
8. Insert the syringe needle in the green PEEK sleeve.
9. Remove the stainless steel nut and ferrule from the other end of the syringe union.
10. Insert the needle and sleeve into the stainless steel nut and ferrule, pushing the sleeve in as far as it will go.
11. Insert the syringe needle, sleeve, nut, and ferrule, into the syringe union, pushing the sleeve in as far as it will
go.
12. Use two 1/4 inch wrenches to tighten the nut.
NanoSpray
®
Ion Source
Operator Guide
85 / 92
RUO-IDV-05-0811-F
Connecting the Syringe Using PEEK-lined Fused Silica Tubing