Figure 4-14 Beam Aimed into Aperture (NanoSpray
®
Interface)
Tip!
When the illuminator is aimed correctly, there is little or no reflection or scattering of the beam on the
curtain plate when the rail is in the operating position.
4. Begin infusing the sample to be analyzed at between 0.5 µL/min and 1.0 µL/min.
CAUTION: Potential System Contamination. Do not insert the end of the emitter tip into
the curtain plate aperture. Make sure that the emitter tip is at least 2 mm to 5 mm
outside the aperture. Spraying too close to the aperture can cause contamination of the
mass spectrometer.
5. Adjust the X-Y-Z positioning unit until the emitter tip is visible on the monitor.
Operator Guide
NanoSpray
®
Ion Source
RUO-IDV-05-0811-F
38 / 92
NanoSpray
®
III Head Assembly and Installation