Figure 4-15 Emitter Tip Position: OptiFlow
TM
Interface (left) and NanoSpray
®
Interface
(right)
Description
Item
Orifice plate
1
Curtain plate
2
Emitter tip
3
2 mm to 5 mm
4
Heater
5
6. Set the
Ion Source Gas 1 (GS1)
and
Ion Source Gas 2 (GS2)
parameters to 0, and the
IonSpray
Voltage (IS)
or
IonSpray Voltage Floating (ISVF)
parameter to 100.
Within a few minutes, a droplet becomes visible at the tip.
7. Increase the IS or ISVF parameter until droplets spray towards the aperture in short bursts.
8. Increase the GS1 parameter until the plume is visible.
show the monitor image when the illuminator is adjusted correctly and the sample
is flowing freely. The position of the emitter tip might vary depending on the application.
Tip!
If the plume seems detached from the emitter tip, then adjust illuminator slightly.
NanoSpray
®
Ion Source
Operator Guide
39 / 92
RUO-IDV-05-0811-F