Figure 4-3 NanoSpray III Head
Description
Item
Sprayer fitting
1
Finger-tight hex nut
2
Small sample line fitting
3
Union thumbscrew
4
High-voltage rail
5
Rail thumbscrew
6
Prepare the Sprayer Fitting
Prepare the sprayer fitting before using the NanoSpray
®
III head for the first time.
Refer also to the
NanoSpray
®
III Head and Emitter Tip Assembly Quick Start Guide
.
CAUTION: Potential System Damage. Be careful when dismantling the NanoSpray
®
III head
because it contains small parts.
1. Remove the sprayer fitting from the NanoSpray head assembly.
Operator Guide
NanoSpray
®
Ion Source
RUO-IDV-05-0811-F
28 / 92
NanoSpray
®
III Head Assembly and Installation