11. Use two 1/4 inch wrenches to tighten the nut.
Tip!
Inject solution and make sure droplets come out of the tubing, to verify the connections.
Figure 4-8 Syringe Union
12. Remove the fitting from the upstream side of the straight union.
13. Insert the fused silica tubing through the tan PEEK sleeve.
14. Cut the end of the fused silica tubing and then clean it with a wipe dampened with methanol or isopropanol.
15. Insert the sleeve and tubing into the PEEK fitting removed in step
, until they protrude about 2 mm from
the tip of the fitting.
NanoSpray
®
Ion Source
Operator Guide
33 / 92
RUO-IDV-05-0811-F
NanoSpray
®
III Head Assembly and Installation