Spray Plume Troubleshooting Tips
Before troubleshooting the spray plume, verify that liquid is arriving at the sprayer head.
Corrective Action
Possible Cause
Symptom
a. Use the X-Y-Z adjustment knobs to move
the sprayer to the correct position.
b. Turn on the illuminator and then verify
that it is operational.
c. Inspect the LC pump pressure. If it is too
high, then replace the emitter tip.
a. The X-Y-Z settings are incorrect.
b. The illuminator is off or faulty.
c. The emitter tip is clogged or damaged.
Inability to view the
spray plume on the
monitor
a. Replace the emitter tip.
a. There is dirt on the emitter tip.
Misshapen or
incorrectly angled
spray plume
a. Adjust the Source/Gas parameters.
a. Source/Gas parameter (GS1, IS, or ISVF)
settings are incorrect.
No spray plume
Monitor and Camera Troubleshooting Tips
Corrective Action
Possible Cause
Symptom
Refer to
or
Incorrectly adjusted camera or illuminator
Poor picture quality
on monitor
Rotate the camera in the mount.
Incorrectly adjusted camera
Inverted image on
monitor
NanoSpray
®
Ion Source
Operator Guide
73 / 92
RUO-IDV-05-0811-F
Troubleshooting