![SCIEX NanoSpray Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/sciex/nanospray/nanospray_operators-manual_1217533065.webp)
5. Clean the aperture in the nano cell heater assembly with a syringe using the cleaning solution. Refer to the
list of
for information about the cleaning solution.
6. Put the nano cell heater assembly on a 100 mL beaker as shown in
Figure 6-11 Nano Cell Heater Assembly on Beaker and Syringe
7. Fill the 5 mL syringe with 5 mL cleaning solution.
8. Inject the cleaning solution through the aperture of the nano cell heater assembly.
9. Repeat step
three times.
10. Wipe the nano cell heater assembly with a lint-free wipe dampened with water.
11. Wipe the nano cell heater assembly with a lint-free wipe dampened with the cleaning solution.
Note:
If more rigorous cleaning is required, then use the brush supplied in the cleaning kit.
12. Wait until the nano cell heater assembly is dry.
13. Inspect the nano cell heater assembly for solvent stains or lint, removing any residue with a clean, slightly
damp, lint-free wipe.
NanoSpray
®
Ion Source
Operator Guide
65 / 92
RUO-IDV-05-0811-F
Ion Source Maintenance