b. Type
RPMT
and then click
Send to instrument
.
The voltage appears below the command.
5. Change the voltage. Type
WRPMTV <XXX>
where <XXX> is the new voltage and then click
Send to
instrument
.
Type a value for the voltage that is based on the raw counts reported in the detector test. Refer to
.
• If the raw counts are greater than 15 000, then enter a voltage that is 5 V less than the reported voltage.
• If the raw counts are less than 15 000, then enter a voltage that 5 V more than the reported voltage.
6. Test the detector. Refer to
• If at least one of the capillaries has a raw count that is approximately 15 000, then no changes are required.
Write down the voltage.
• If there are no capillaries with a raw count that is approximately 15 000, then repeat step
until the raw counts in at least one of the capillaries is approximately 15 000. Write down the voltage.
Test the Detector
Use this test to make sure that all of the capillaries in the C100
HT
Biologics Analyzer system work correctly.
1. In the C100
HT
software, click
to go to Service mode.
2. In
Service
mode, click the
System Check
tab.
3. Click
Detector Test
and then click
Start
.