CAUTION: Potential Wrong Result. Do not click
(Maximize) to maximize any of the
windows in the following step. If any of the windows are maximized, then the title bar
buttons for windows in the DataReviewer software might not be visible.
a. If the 32 Karat software window is maximized, click
(Restore).
b. Click
File > Open
and open a data file.
c. Drag the edges of the data window so that it fills the 32 Karat software window as shown in the following
figure.
9. Click
File > Exit
to close the 32 Karat software.
10. In the
32 Karat Software Enterprise
window, click
File > Exit
.
11. Put the customer reference DVD in the DVD drive on the controller and then copy the
C100
HT
Biologics
Analyzer Operator Guide
to the Windows desktop.
Set the Instrument ID
1. Write down the serial number found in the
SN
field in the nameplate on the back of the C100
HT
Biologics
Analyzer system.
2. Start the C100
HT
software and type
Service
in the
USER ID
field,
ABsciexService1
in the
Password
field, and then click
OK
.
Installation Guide
C100
HT
Biologics Analyzer System
RUO-IDV-09-3829-A
14 / 31
Installation Instructions