The Cartridge Status panel in the C100
HT
software updates to display the cartridge type, ID, remaining runs,
remaining normalization runs, status, and other information.
Verify the Alignment of the Transport System
Required Materials
• 96-well plate
• Empty buffer plate
WARNING! Personal Injury Hazard. Keep fingers clear of the buffer plate carrier and
the sample door. When the sample door is open for more than five minutes, the
tray automatically moves to the park position.
CAUTION: Potential System Damage. Move the transport system one step at a time when
performing the alignment. If the vertical or horizontal position is changed too quickly, the
buffer plate or 96-well sample plate might collide with the capillary tips and damage them.
1. Log into the C100
HT
software as the service user. Type
Service
in the
USER ID
field,
ABsciexService1
in the
Password
field, and then click
OK
.
2. In the C100
HT
software, click
to go to Service mode.
Installation Guide
C100
HT
Biologics Analyzer System
RUO-IDV-09-3829-A
16 / 31
Installation Instructions