7. Click
Send to instrument
.
The injection begins.
8. When the sample injection is complete, test the detector. Refer to
be between 14 000 and 15 000.
• If at least one of the capillaries has a raw count that is approximately 15 000, then no changes are required.
• If none of the capillaries has a raw count that is approximately 15 000, then set the photomultiplier voltage.
Set the Photomultiplier Voltage
.
Set the Photomultiplier Voltage
1. Start the C100
HT
software and type
Service
in the
USER ID
field,
ABsciexService1
in the
Password
field, and then click
OK
.
2. In the C100
HT
software, click
to go to Service mode.
3. In
Service
mode, click the
Terminal
tab.
4. Check the voltage.
a. Select
Direct Commands
.
C100
HT
Biologics Analyzer System
Installation Guide
19 / 31
RUO-IDV-09-3829-A
Installation Instructions